JJF 2425-2026 for carbon dioxide incubators and JJF 2426-2026 for pneumatic ballistic extracorporeal pressure wave therapy devices take effect on 25 December 2026. The first sets nine calibration points across three parameters; the second covers output intensity, penetration depth and collision frequency.
On 26 November 2026 the State Administration for Market Regulation (SAMR) announced the approval of two national metrology technical specifications: JJF 2425-2026, Calibration specification for carbon dioxide incubators, and JJF 2426-2026, Calibration specification for pneumatic ballistic extracorporeal pressure wave therapy devices. Both take effect on 25 December 2026. SAMR says the documents fill gaps in calibration rules for life science laboratories and rehabilitation equipment and provide a common metrological basis for medical institutions and testing bodies.
Carbon dioxide incubators simulate the internal biological environment for cell and tissue culture and are widely used in immunology, genetics and pharmacology; their metrological performance directly affects culture quality and diagnostic results. JJF 2425-2026 covers temperature, carbon dioxide concentration and relative humidity, arranging nine calibration points across upper, middle and lower levels and assessing indication error, uniformity and fluctuation. The default points are 37.0 degrees Celsius, 5.0 % carbon dioxide and 95 % RH, while commonly used set points may be selected to reflect actual working conditions, balancing consistency with practical operability.
Pneumatic ballistic extracorporeal pressure wave therapy devices act directly on the human body in rehabilitation, orthopaedics, pain management and traditional Chinese medicine therapy, so the accuracy of output parameters bears on patient safety. JJF 2426-2026 addresses three indicators: output intensity, assessed through the accuracy and stability of pressure wave output energy; penetration depth, assessed through the effective depth of action in tissue; and collision frequency, assessed through the rhythm of pressure wave application. The specification also states a scope boundary, excluding pneumatic ballistic lithotripters, which laboratories must check before accepting work.
The two specifications serve a dual purpose. For manufacturers they set a benchmark for design against clinical use and for factory performance control; for medical institutions and testing bodies they provide a common rule that makes calibration results comparable and traceable. SAMR plans organisation and training sessions and further specifications in medical metrology.
Four practical consequences follow for laboratories. First, scopes of accreditation or qualification need updating: a laboratory adding either specification must complete method validation, measurement uncertainty evaluation, verification of the traceability chain for reference instruments and authorisation of personnel. Second, carbon dioxide incubators are no longer assessed at a single point; uniformity and fluctuation must be recorded at upper, middle and lower positions, requiring revised record forms and data-processing procedures. Third, energy and frequency measurements on pressure wave devices involve dedicated loading and safety precautions, so working instructions and safety plans should be reviewed, with strict separation between therapy devices and lithotripters to avoid out-of-scope certificates. Fourth, with application starting on 25 December 2026, hospitals, disease control institutions and third-party laboratories should inventory equipment and schedule calibration ahead of the year-end rush. The official texts of the specifications prevail.